PULSE 6 SERVICE MANUAL
PAGE 4.3
SEPT 2011SUNRISE MEDICAL
12. If the above tests pass, replace the wiring
harness leading from the batteries through the
circuit breaker and control module. Retest as
necessary.
.
10. If battery voltage is not present, use a
Multimeter and measure for continuity across the
fusible links connected to positive (+) terminal of
each battery (fi g. 4.7). Normal resistance is less
than 1 ohm. If open, replace the defective battery
harness. Retest as necessary.
.
Caution: To prevent damage to the Multimeter,
ensure that the batteries are disconnected.
11. For chairs prior to PLS-100808, Check for
continuity across the circuit breaker (fi gure 4.3.3).
If Normal resistance is less than 1 ohm. If open,
replace the circuit breaker. Retest as necessary.
Note: Access to the circuit breaker terminals is
easier if the front battery is removed..
9. Verify that battery voltage is present at each
connector leading to the batteries (fi g. 4.6).
fig. 4.3.1
fig. 4.3.2
Section 4
Rnet Troubleshooting Procedures (cont.)
fig. 4.3.3