PULSE 6 SERVICE MANUAL
PAGE 6.1
SEPT 2011SUNRISE MEDICAL
Note: This procedure may be done with or without removing the motor from the power chair. If
the motor/gearbox is to be removed from the power chair, refer to Motor and Gearbox Removal,
(Section7) of this manual. Consider replacing any motor/gearbox assembly that does not pass
the following test.
1. Inspect the output shafts for signs of grease
leakage ( gure. 6.1.1).
2. Inspect the cotter pin connecting the gearbox
release arm to the motor/gearbox assembly.
Ensure that it is tight and not damaged ( gure.
6.1.2). Chairs after serial number PLS-101619
and with s/n pre x PLS6A & PLS6B, did not
utilize this feature.
3. Manually operate each gearbox release lever
to ensure that they operate easily and fully engage
and disengage
Note: When removing the brush assemblies,
remove one brush assembly at a time. Note its
orientation so that it can be inserted into the
brush holder the same way it was removed.
Caution: Use care not to damage the brush
cap with excessive force in the next step.
Motor/Gearbox Inspection
Section 6
fig. 6.1.1
fig. 6.1.2
fig. 6.1.3
4. Remove the brush by using a common
screwdriver to unscrew the cap ( gure. 6.1.3).