PULSE 6 SERVICE MANUAL
PAGE 0.2
2009SUNRISE MEDICAL
When setting up the components of the chair, complete the following checklist to ensure proper and
safe operation of the equipment.
Check :
Are the batteries fully charged?
Test battery voltage with D.C. meter across the terminals of batteries. The measurement should a.
be above 12 volts D.C.
If not, fully charge the batteries.b.
Are all necessary power components installed and connected ?
Input device (normally Joystick) a.
Cable from Joystick to the Bus Line b.
Control Module; for the Pulse located at back of chair behind shroud c.
Are all necessary connections fastened or inserted?
Battery connectors to the batteries a.
Cable between Joystick and the Control b.
Both Motor Connectors to the Control Module.c.
Is the Drive Gear engaged?
With the power off the chair should not move if pushed from behind. a.
If the chair moves when pushed, refer to proper operation of Drive Gear Engagement section.b.
Does Display light up when Power On/Off switch is depressed?
If no - recheck the 4 checks listed above then refer to the section on Diagnostics.a.
If yes – the Power Wheelchair is ready to driveb.
Basic Setup