PULSE 6 SERVICE MANUAL
PAGE 7.15
2009SUNRISE MEDICAL
Section 7
Caution: the actuator and micro-switch assembly wires are fragile. Use caution when removing
the actuator and micro-switch assembly.
8. Remove the actuator and micro-switch
assembly as one part if the entire assembly is to
be replaced ( g. 7.44). Reverse the procedures
to install and retest as necessary.
Note: Perform this step if only the actuator is
to be replaced.
9. Note the connections of the wires and
separate them ( g. 7.45). Remove the actuator.
10. Reverse the previous procedure to install
the actuator/micro-switch assembly. Retest as
necessary.
Removal Procedures (cont.)
fig. 7.44
fig. 7.45