PULSE 6 SERVICE MANUAL
PAGE 6.1
2009SUNRISE MEDICAL
Note: This procedure may be done with or without removing the motor from the power chair.
If the motor/gearbox is to be removed from the power chair, refer to the Motor and Gearbox
Removal section of this manual.
Consider replacing any motor/gearbox assembly that does not pass the following test. Refer
to the Motor Gearbox Removal section of this manual.
1. Inspect the output shafts for signs of grease
leakage ( g. 6.1).
2. Inspect the cotter pin connecting the
gearbox release arm to the motor/gearbox
assembly. Ensure that it is tight and not damaged
( g. 6.2)
3. Manually operate each gearbox release
lever to ensure that they operate easily and fully
engage and disengage
Note: When removing the brush assemblies,
remove one brush assembly at a time. Note its
orientation so that it can be inserted into the
brush holder the same way it was removed.
Caution: Use care not to damage the brush
cap with excessive force in the next step.
4. Remove the brush by using a common
screwdriver to unscrew the cap ( g. 6.1).
Motor/Gearbox Inspection
Section 6
fig. 6.1
fig. 6.2
fig. 6.3