PULSE 6 SERVICE MANUAL
PAGE 7.4
2009SUNRISE MEDICAL
Section 7
6. Remove the gearbox release lever from
the motor/gearbox assembly by removing the
cotter pin ( g. 7.12). Use a new cotter pin when
replacing the lever.
Note: For the next procedure, use either a
ball-type 5 mm Allen wrench or a cut-down 5
mm Allen wrench.
7. Remove the 4 bolts securing the motor/
gearbox assembly to the motor mount using a 5
mm Allen wrench ( g. 7.13).
8. Lift the motor out from the side of the power
chair ( g. 7.14).
9. Replace the motor/gearbox assembly by
reversing the previous procedures.
Removal Procedures (cont.)
fig. 7.12
fig. 7.13
fig. 7.14